• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH VIPER SYSTEM FENESTRATED CORTICAL FIX POLYAXIAL SCREW 5.5 X 6 X 45MM; ORTHOSIS, SPINAL PEDICLE FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH VIPER SYSTEM FENESTRATED CORTICAL FIX POLYAXIAL SCREW 5.5 X 6 X 45MM; ORTHOSIS, SPINAL PEDICLE FIXATION Back to Search Results
Model Number 186727645
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/16/2022
Event Type  malfunction  
Event Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2022, both the setscrew and the screw tulip were damaged during the final locking process.Both implants had to be replaced.There was a fifteen (15) minute surgical delay.There were fragments generated and the fragments were removed.The procedure was completed successfully.There was no patient consequence.This report involves one (1) screw.This is report 1 of 2 for (b)(4).
 
Manufacturer Narrative
Additional narrative: additional device product codes: pml.Complainant part is not expected to be returned for manufacturer review/investigation.Reporter is a j&j employee.The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the photo.Visual analysis of the photo revealed that the tulip of the mis ti cfx fen poly 6x45 appears to be deformed.No indication suggesting the tulip broke or detached was observed in the provided images.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for mis ti cfx fen poly 6x45.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIPER SYSTEM FENESTRATED CORTICAL FIX POLYAXIAL SCREW 5.5 X 6 X 45MM
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle
SZ  
Manufacturer Contact
kara ditty-bovard
chemin-blanc 38
le locle 02400
SZ   02400
6103142063
MDR Report Key13805034
MDR Text Key288037383
Report Number1526439-2022-00410
Device Sequence Number1
Product Code MNI
UDI-Device Identifier10705034282565
UDI-Public(01)10705034282565
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K160879
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number186727645
Device Catalogue Number186727645
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/17/2022
Initial Date FDA Received03/18/2022
Supplement Dates Manufacturer Received04/18/2022
Supplement Dates FDA Received04/18/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
SINGLE-INNER SETSCREW
-
-