• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION RESOLUTION CLIP; CLIP, HEMOSTATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION RESOLUTION CLIP; CLIP, HEMOSTATIC Back to Search Results
Model Number M00522610
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/18/2022
Event Type  malfunction  
Manufacturer Narrative
Initial reporter address: (b)(6).Medical device code: (b)(4).The device has been received for analysis; however, the analysis has not yet been completed.Upon completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation that a resolution clip device was used in the colon during an endoscopic mucosal resection (emr) procedure performed on february (b)(6) 2022.During the procedure, the clip was able to grasp and lock onto tissue; however, the clip was unable to release from the catheter to deploy.The procedure was completed with another resolution clip device.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Manufacturer Narrative
E1: initial reporter address: (b)(6).H6: medical device code a15 captures the reportable event of clip unable to release from the catheter.H10: investigation results the returned resolution clip device was analyzed, and a visual evaluation noted that the device was returned with the clip assembly attached.The outer-sheath was not returned with the device.No problems were identified during microscopic examination.Functional evaluation was performed and the device can open its arms within specification.The clip assembly was able to perform the first activation and could be released from the catheter without problem.No problems with the device were noted.The reported event was not confirmed.No problems were found that could have reduced the performance of the device during the procedure.The returned device review showed no evidence of either the alleged(s) or any defect which could have contributed to the event.Therefore, the most probable root cause for this complaint is no problem detected.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
 
Event Description
It was reported to boston scientific corporation that a resolution clip device was used in the colon during an endoscopic mucosal resection (emr) procedure performed on february 18, 2022.During the procedure, the clip was able to grasp and lock onto tissue; however, the clip was unable to release from the catheter to deploy.The procedure was completed with another resolution clip device.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTION CLIP
Type of Device
CLIP, HEMOSTATIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
2546 calle primera
propark, coyol
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13805599
MDR Text Key287432500
Report Number3005099803-2022-01433
Device Sequence Number1
Product Code MCH
UDI-Device Identifier08714729504795
UDI-Public08714729504795
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K040148
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00522610
Device Catalogue Number2261
Device Lot Number0028229162
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/21/2022
Initial Date FDA Received03/18/2022
Supplement Dates Manufacturer Received03/24/2022
Supplement Dates FDA Received04/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/19/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age66 YR
Patient SexMale
Patient Weight66 KG
-
-