• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 TITANIUM LOW-PROFILE IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 TITANIUM LOW-PROFILE IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Catalog Number 0606150J
Device Problems Degraded (1153); Fracture (1260); Peeled/Delaminated (1454)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/04/2022
Event Type  malfunction  
Event Description
It was reported that approximately seven years post port placement procedure via the internal jugular vein for chemotherapy and when the port system was removed, it was allegedly found that the catheter covered the subcutaneous tissue like a sheath.It was further reported that when the catheter was removed, the part of the surface of the catheter peeled off and remained subcutaneous.The port system was not used for approximately three years and was not flushed.There was no reported patient injury.
 
Manufacturer Narrative
The catalog number identified has not been cleared in the us but is similar to the titanium low-profile implantable port, groshong single-lumen, 8f that are cleared in the us.The pro code and 510 k number for the titanium low-profile implantable port, groshong single-lumen, 8f are identified.As the lot number for the device was not provided, a review of the device history records could not be performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the titanium low-profile implantable port, groshong single-lumen, 8f that are cleared in the us.The pro code and 510 k number for the titanium low-profile implantable port, groshong single-lumen, 8f are identified in d2 and g4.H10: manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: one bard ti/lp implantable port attached to a groshong catheter was returned for evaluation.Gross visual, microscopic visual and functional evaluation were performed on the returned device.The investigation is confirmed for the reported peeled and identified degraded issues as the catheter was noted to be degraded from the 7.0cm depth mark to the 12.0cm depth mark and the catheter material appeared to be peeled off in the same region.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that approximately seven years post port placement procedure via the internal jugular vein for chemotherapy and when the port system was removed, it was allegedly found that the catheter covered the subcutaneous tissue like a sheath.It was further reported that when the catheter was removed, the part of the surface of the catheter peeled off and remained subcutaneous.The port system was not used for approximately three years and was not flushed.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TITANIUM LOW-PROFILE IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, 8F
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key13819973
MDR Text Key287524437
Report Number3006260740-2022-00870
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741098253
UDI-Public(01)00801741098253
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K050310
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0606150J
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/09/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/21/2022
Initial Date FDA Received03/19/2022
Supplement Dates Manufacturer Received05/02/2022
Supplement Dates FDA Received05/05/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
-
-