• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, LLC RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, LLC RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60 VENTILATOR,JAPAN OPT: CFLEX,AVAPS
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/27/2022
Event Type  malfunction  
Manufacturer Narrative
Reporting institution phone number - (b)(6).Reporter phone number - (b)(6).
 
Event Description
It was reported that there was light leaking from the lower left part of the screen.There was no patient involvement at the time the issue was discovered.The service engineer confirmed that light was leaking from the lower left part of the screen.The service engineer replaced the lcd, and the issue was resolved.
 
Manufacturer Narrative
The lcd, user interface assembly was returned to the manufacturer for failure analysis.The returned component failed all testing, the burnt-out cold cathode fluorescent lamp (ccfl) backlight causes the dim display and leaking light.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer Contact
melissa abbott
2271 cosmos court
carlsbad, CA 92011
7609187300
MDR Report Key13830011
MDR Text Key287694331
Report Number2031642-2022-00709
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60 VENTILATOR,JAPAN OPT: CFLEX,AVAPS
Device Catalogue Number1053618
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/27/2022
Initial Date FDA Received03/20/2022
Supplement Dates Manufacturer Received06/03/2022
Supplement Dates FDA Received06/27/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/19/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-