• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TWINFIX ULTRA PK 4.5MM W/2 UB -WHT &BL; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TWINFIX ULTRA PK 4.5MM W/2 UB -WHT &BL; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72202595
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Joint Dislocation (2374); Impaired Healing (2378); Skin Disorders (4543); Swelling/ Edema (4577)
Event Date 02/08/2022
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: case-(b)(4).
 
Event Description
It was reported that a patient underwent microscopic examination of the right knee, reconstruction of the medial patellofemoral ligament and release of the lateral retinaculum due to the dislocation of the right patella.After the procedure using twinfix anchor, it was found that the wound of the right knee exuded repeatedly, and the dressing was changed for many times but there was no effect.A debridement was performed due to exudation.The patient current status is fine.
 
Manufacturer Narrative
H10: internal complaint reference: (b)(4).H3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.There was no way to determine if the device contributed to the reported event.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that there was insufficient information to tie the reported complaint to specific line items within the risk file.Based on the information provided, a debridement was performed due to exudation and the patient is currently doing fine.Therefore, no further clinical/medical assessment is warranted at this time.Should any additional relevant patient information be provided, this case would be re-assessed.The complaint was not confirmed.Factors that could have contributed to the reported event include an adverse biological response to the procedure, inadequate cleaning of the surgical site, or failure of a concomitant device.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TWINFIX ULTRA PK 4.5MM W/2 UB -WHT &BL
Type of Device
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13835612
MDR Text Key287530599
Report Number1219602-2022-00394
Device Sequence Number1
Product Code MBI
UDI-Device Identifier03596010647467
UDI-Public03596010647467
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K093228
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Consumer,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number72202595
Device Catalogue Number72202595
Device Lot Number2056323
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/28/2022
Initial Date FDA Received03/20/2022
Supplement Dates Manufacturer Received04/06/2022
Supplement Dates FDA Received04/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/12/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age13 YR
Patient SexFemale
-
-