DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Electrical /Electronic Property Problem (1198)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/22/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Communications/ interviews: the getinge field service engineer (fse) indicated that during a phone conversation, he asked them to check if the console was properly engaged in the cart.They were able to pull out the console from the pump and push it back in, but over the fse never heard the console engage the cart and believed they did not push hard enough to reengage.The fse attempted to have them try again, but they were anxious to place the patient on a new pump and did that.The first pump was being sent to biomed.At this time, the customer has not requested getinge to evaluate the iabp.Additional information is being requested from the customer with regard to the repair and status of the iabp.A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) gave a battery in use message while connected to outlet power.The unit was swapped out with another to continue therapy.No patient harm, serious injury or adverse event was reported.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
E3 is unknown (initially it is submitted as "nurse").Additional information: e1(event site telephone: (b)(6), event site email: (b)(6).
|
|