• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/22/2022
Event Type  malfunction  
Manufacturer Narrative
Communications/ interviews: the getinge field service engineer (fse) indicated that during a phone conversation, he asked them to check if the console was properly engaged in the cart.They were able to pull out the console from the pump and push it back in, but over the fse never heard the console engage the cart and believed they did not push hard enough to reengage.The fse attempted to have them try again, but they were anxious to place the patient on a new pump and did that.The first pump was being sent to biomed.At this time, the customer has not requested getinge to evaluate the iabp.Additional information is being requested from the customer with regard to the repair and status of the iabp.A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) gave a battery in use message while connected to outlet power.The unit was swapped out with another to continue therapy.No patient harm, serious injury or adverse event was reported.
 
Event Description
N/a.
 
Manufacturer Narrative
E3 is unknown (initially it is submitted as "nurse").Additional information: e1(event site telephone: (b)(6), event site email: (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key13839348
MDR Text Key287562162
Report Number2249723-2022-00579
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 02/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/22/2022
Initial Date FDA Received03/21/2022
Supplement Dates Manufacturer Received02/20/2024
Supplement Dates FDA Received02/20/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/17/2020
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
IAB CATHETER
Patient Age92 YR
Patient SexFemale
Patient Weight68 KG
-
-