The customer provided instrument log files for further investigation.Based on the instrument log review, the root cause for the suspected discrepant low sodium result when compared to repeated na+ results generated on the lab instrument was due to an unstable na+ sensor within the measurement cartridge.The na+ sensor was exposed to a quaternary ammonium compound (qac) such as benzalkonium or chlorhexidine, noted as a "sodium sensor interferent detected'' error.The customer stated that they use cleaning wipes that contain a qac.The customer has replaced the measurement cartridge and has been instructed to cease using disinfectants containing qacs.The customer is operational.
|