• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. RAPIDPOINT 500E BLOOD GAS SYSTEM; RP 500E

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. RAPIDPOINT 500E BLOOD GAS SYSTEM; RP 500E Back to Search Results
Catalog Number 11416755
Device Problem Low Test Results (2458)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2022
Event Type  malfunction  
Manufacturer Narrative
The customer provided instrument log files for further investigation.Based on the instrument log review, the root cause for the suspected discrepant low sodium result when compared to repeated na+ results generated on the lab instrument was due to an unstable na+ sensor within the measurement cartridge.The na+ sensor was exposed to a quaternary ammonium compound (qac) such as benzalkonium or chlorhexidine, noted as a "sodium sensor interferent detected'' error.The customer stated that they use cleaning wipes that contain a qac.The customer has replaced the measurement cartridge and has been instructed to cease using disinfectants containing qacs.The customer is operational.
 
Event Description
The customer reported a discrepant low sodium result from their rp500e instrument compared to repeat testing of the same sample on a non-siemens lab instrument.There is no report of injury due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDPOINT 500E BLOOD GAS SYSTEM
Type of Device
RP 500E
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
northern road
chilton industrial estate
sudbury, CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
tim krauskopf
2 edgewater drive
norwood, MA 02062
6107052212
MDR Report Key13860089
MDR Text Key287672150
Report Number3002637618-2022-00020
Device Sequence Number1
Product Code CHL
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K192240
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number11416755
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/07/2022
Initial Date FDA Received03/23/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-