Model Number 777F8
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/21/2022
Event Type  malfunction  
Event Description
It was reported that the balloon of swan-ganz cco catheter did not inflate during balloon testing before use.There was no patient involvement.
 
Manufacturer Narrative
The reported event of did not inflate was confirmed.Balloon did not inflate due to a tear, approximately 0.01 inches in length near the central area of the balloon.The edges of latex did not appear to match at the torn location.All through lumens were patent without any leakage or occlusion.No other visible damage or abnormality was observed from catheter body.Further evaluation regarding supplier related quality issues is under investigation.A device history record review was completed and documented that device met all specifications upon distribution.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN GANZ CCOMBO
Type of Device
CATHETER, OXIMETER, FIBER-OPTIC
Manufacturer (Section D)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco, puerto rico 00610
*  00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco, puerto rico 00610
*   00610
Manufacturer Contact
jessica atallah
1 edwards way
irvine, CA 92614
MDR Report Key13860472
MDR Text Key287673707
Report Number2015691-2022-04558
Device Sequence Number1
Product Code DQE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K040287
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date03/29/2023
Device Model Number777F8
Device Catalogue Number777F8
Device Lot Number63707542
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/17/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/03/2022
Initial Date FDA Received03/23/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/02/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-