ZIMMER BIOMET, INC. ARCOS CON SZ A HI 50MM TRL; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Difficult or Delayed Positioning (1157)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/25/2022
Event Type  malfunction  
Event Description
It was reported the screw connecting the proximal body and distal stem had a hard time tightening and loosening.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
(b)(4).Concomitant medical products: 11-300816- arcos 16x150mm spl tpr dist- 943650.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2022 - 00649.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No additional information on the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCOS CON SZ A HI 50MM TRL
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13868957
MDR Text Key287826214
Report Number0001825034-2022-00648
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090757
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number31-301310
Device Lot Number427427
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/28/2022
Initial Date FDA Received03/23/2022
Supplement Dates Manufacturer Received04/04/2022
Supplement Dates FDA Received04/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/12/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberNI
Patient Sequence Number1
Treatment
SEE H10 NARRATIVE.
Patient SexMale
-
-