Manufacturer's investigation conclusion: incidental findings: damaged front overlay.The reported event of a motor stop was confirmed via the log file; however, the reported event of a loose cable between the console and motor was not confirmed.The centrimag 2nd generation primary console (serial #: (b)(6)) was returned for analysis and a log file was downloaded for review with events spanning approximately 3 days (17feb2022 ¿ 18feb2022, 23mar2022 per time stamp).The events occurring on 23mar2022 took place during lab testing at abbott.On (b)(6) 2022 at 14:01, a sub-fault ¿sf_ifd_shutdown_detected¿ activated and was followed by ¿motor disconnected: m2¿ and ¿flow below minimum: f3¿ alarms.The speed drop to 0 rpm with a flow slightly above 0 lpm.The alarms were able to be muted and cleared and the speed and flow recovered a couple minutes later.The console was powered off shortly after at 14:17.There were no other notable events active in the log file.The centrimag 2nd generation primary console was returned for analysis to the service depot and the reported event was unable to be reproduced.The console was tested with the returned and associated motor and was run for several days on a test loop.No alarms were observed.A full functional test was performed, and the unit passed all tests.The root cause for the reported event was unable to be conclusively determined through this analysis.The device history records were reviewed for the centrimag 2nd generation primary console (serial #: (b)(6)) and the console was found to pass all manufacturing and qa specifications.The 2nd generation centrimag system operating manual section 4 entitled "warnings & precautions" warns "one additional 2nd generation centrimag primary console, motor and flow probe are required as backup system in the immediate vicinity of each patient whenever the centrimag or pedivas blood pump is used.The backup console must be connected to the backup motor and to the backup flow probe, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the main console, motor or flow probe experience a malfunction.".The 2nd generation centrimag system operating manual section 10 entitled "emergency and troubleshooting" states that "the recommended practice whenever there is a 2nd generation centrimag primary console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the backup motor and console to continue patient support.Do not exchange individual motors or individual consoles during patient support.".No further information was provided.The manufacturer is closing the file on this event.
|