• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® I.C. FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® I.C. FOLEY CATHETER Back to Search Results
Model Number 0165SI18
Device Problem Partial Blockage (1065)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/28/2022
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event Description
It was reported that the foley catheters kept clogging on patient.
 
Event Description
It was reported that the foley catheters kept clogging on patient.
 
Manufacturer Narrative
The reported event was inconclusive because no sample was returned.It is unknown whether the device had met relevant specifications.The product was used for urological care.It was unknown whether the product had caused the reported failure.The potential root cause for this failure could be due to wrong duration setting on air purging/plc failure.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: ''warning: on catheter, do not use ointments or lubricants having a petrolatum base.They will damage the catheter and may cause balloon to burst.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practices and applicable local, state and federal laws and regulations.Visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.Please consult product label and insert for any indications, contraindications, hazards, warnings, cautions and directions for use.Proper techniques for urinary catheter maintenance 1) secure the foley catheter.Use the statlock® foley stabilization device if provided.2) maintain a closed drainage system by utilizing pre-connected, sealed catheter-tubing junctions.3) maintain unobstructed urine flow and keep the catheter and collection tube free from kinking.4) keep the collection bag below the level of the bladder or hips at all times.5) empty the collection bag regularly using a separate, clean collection container for each.Foley catheter removal 1) to deflate catheter balloon: gently insert a luer lock or slip tip syringe in the catheter valve.Never use more force than is required tomake the syringe ¿stick¿ in the valve 2) allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently 3) use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation 4) if the balloon will not deflate and if permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol 5) should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.'' the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® I.C. FOLEY CATHETER
Type of Device
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key13869930
MDR Text Key287810273
Report Number1018233-2022-01511
Device Sequence Number1
Product Code MJC
UDI-Device Identifier00801741016769
UDI-Public(01)00801741016769
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K910318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number0165SI18
Device Catalogue Number0165SI18
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/28/2022
Initial Date FDA Received03/23/2022
Supplement Dates Manufacturer Received07/15/2022
Supplement Dates FDA Received07/31/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexMale
-
-