• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. DUAL CHAMBER TEMPORARY PACEMAKER; PULSE-GENERATOR, PACEMAKER, EXTERNAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. DUAL CHAMBER TEMPORARY PACEMAKER; PULSE-GENERATOR, PACEMAKER, EXTERNAL Back to Search Results
Model Number 5392
Device Problems Display or Visual Feedback Problem (1184); Battery Problem (2885)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/18/2022
Event Type  malfunction  
Event Description
The external pulse generator (epg) which returned into service subsequently tested out of specification during manufacturer's analysis.There was no patient involvement.
 
Manufacturer Narrative
Product analysis: analysis was able to confirm the customers comment that external pulse generator (epg) displayed an error code.It was noted an error occurred, the main printed circuit board (pcb) board was replaced.The battery shorting bar was contaminated.One case screw was contaminated.All found defective parts were replaced and all other identified issues were resolved.The device then passed all functionality tests.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUAL CHAMBER TEMPORARY PACEMAKER
Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key13902566
MDR Text Key287902916
Report Number2182208-2022-00911
Device Sequence Number1
Product Code DTE
UDI-Device Identifier00643169725126
UDI-Public00643169725126
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2022
Initial Date Manufacturer Received 03/07/2022
Initial Date FDA Received03/25/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/22/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-