• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KAVO DENTAL GMBH MASTERTORQUE LUX M9000 L; DENTAL HANDPIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KAVO DENTAL GMBH MASTERTORQUE LUX M9000 L; DENTAL HANDPIECE Back to Search Results
Model Number M9000 L
Device Problem Dent in Material (2526)
Patient Problem Burn(s) (1757)
Event Date 02/10/2022
Event Type  malfunction  
Event Description
Following information has been supplied by dental office: when patient received a new crown and dentist wanted to polish it after inserting the used dental handpiece heated up and caused a burn on inner cheek.The patient recognized the heat from handpiece.There was no medical care necessary, and burn healed already.
 
Manufacturer Narrative
The analysis prior to the repair showed that the back cap had a strong dent which caused the push button to stick in.This caused unusual inner friction and hence heat up of the push button.The dent is clearly the result of a strong external force, such as a drop.To avoid such issues the user instruction contains already several notes, warnings and requests how to prepare the handpiece for each treatment and how to use it: warning: hazards for the care provider and the patient.In the case of damage, irregular running noise, excessive vibration, untypical warming or when the cutter or grinder cannot be held.Do not use further and notify service.Technical condition: a damaged device or components could injure patients, users and third parties.Only operate devices or components if they are undamaged on the outside.Check that the device is working properly and is in satisfactory condition before each use.If there are any sites of breakage or clearly recognised changes on the surface, the parts must be checked by the service.Safety checks may only be performed by trained service personnel.If the following defects occur, stop working and have the service personnel carry out repair work: malfunctions, damage, irregular noise, excessive vibration, untypical heating, and no firm seating of the grinder or cutter in the instrument.Observe the following instructions in order to guarantee optimum functioning and prevent material damage: service the medical device with care products and systems regularly as described in the instructions for use.Kavo recommends specifying in-house service intervals where the medical device is brought to a professional shop for cleaning, servicing and a function check.Define the service interval depending on the frequency of use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MASTERTORQUE LUX M9000 L
Type of Device
DENTAL HANDPIECE
Manufacturer (Section D)
KAVO DENTAL GMBH
bismarckring 39
biberach / riss, 88400
GM  88400
Manufacturer (Section G)
KAVO DENTAL GMBH
bismarckring 39
biberach / riss, 88400
GM   88400
Manufacturer Contact
klaus reisenauer
bismarckring 39
biberach / riss, 88400
GM   88400
MDR Report Key13937531
MDR Text Key289777148
Report Number3003637274-2022-00016
Device Sequence Number1
Product Code EFB
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
K130560
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Dentist
Type of Report Initial
Report Date 03/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM9000 L
Device Catalogue Number1.007.9900
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/04/2022
Initial Date FDA Received03/29/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/14/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-