|
Model Number 5392 |
Device Problems
Battery Problem (2885); Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/03/2022 |
Event Type
malfunction
|
Event Description
|
The external pulse generator (epg) which was returned into service subsequently tested out of specification during manufacturer¿s analysis.There was no patient involvement.
|
|
Manufacturer Narrative
|
Product analysis: analysis confirmed the customer comment that the output connector assembly was broken.The hanger assembly was broken.The device failed incoming functional test "ventricular pace pulse noise".Main printed circuit board (pcb) was out of specification electrical.The encoder stems were contaminated, but functional.Two knobs were contaminated.The battery drawer sticks (during initial inspection).All found defective parts were replaced and all other identified issues were resolved.The device then passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Further analysis was performed on the external pulse generators (epg) printed circuit board (pcb), for atrial and ventricle pace pulse noise.Analysis found that the atrial and ventricle pace pulse noise measured within test specification requirements.No defect was found.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Fdc code updated.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|