The suspect device was returned to olympus for evaluation.Inspection and testing confirmed the reported event.The forceps elevator was damaged which caused rotation of the knob lever to be rough.In addition, the angulation was reduced and the angulation control knob was loose due to stretched angle wires, the adhesive on the bending section was damaged due to stress, there was evidence of water infiltration, and the switch, cord, connector, and control section were damaged.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the reported event is unable to be determined.However, the cause of the event is likely due to the stroke of the scope knob wire (k-wire) for inverting the forceps elevator is temporarily insufficient due to the cut length of the scope k wire is temporarily shortened with respect to the set value and forceps elevator did not completely return.There is a possibility that the movement (operation) of the scope k wire was temporarily hindered by the influence of moisture (moisture) that entered the scope, or by foreign material adhering and the forceps elevator could not return completely.It is likely that water invaded from the venting connector of the scope connector unit or debris from the venting connector and other foreign material temporarily entered between the check valve and water invaded without closing completely.Olympus will continue to monitor field performance for this device.
|