Additional suspect medical device components involved in the event: product family: dbs-extension, upn: m365nm3138550 , model: nm-3138-55, serial: (b)(4), batch: 7084237.Product family: dbs-extension, upn: m365nm3138550 , model: nm-3138-55, serial: (b)(4), batch: 7085390.Product family: dbs-lead fixation, upn: m365db4600c0 , model: db-4600c, serial: n/a, batch: 26729878.Product family: dbs-lead fixation, upn: m365db4600c0 , model: db-4600c, serial: n/a, batch: 26793896.Product family: dbs-lead fixation, upn: m365db4605c0, model: db-4605c, serial: n/a, batch: 26515106.Product family: dbs-linear leads, upn: m365db2202450 , model: db-2202-45, serial: (b)(4), batch: 7080528.Product family: dbs-linear leads, upn: m365db2202450 , model: db-2202-45, serial: (b)(4), batch: 7080583.
|
Db-1216, (b)(6) the returned ipg passed the visual inspection and revealed no anomalies.Db-2202-45, (b)(6) visual inspection revealed that the lead was cleanly cut into two pieces from the distal end of the lead.The clean cut damage is a result of a typical explant procedure, and it is not considered a failure.No other anomalies were identified on the returned portion of the lead.
|