• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH 7.3MM CANNULATED CONICAL SCREW PARTIALLY THREADED/110MM; PIN, FIXATION, SMOOTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH 7.3MM CANNULATED CONICAL SCREW PARTIALLY THREADED/110MM; PIN, FIXATION, SMOOTH Back to Search Results
Catalog Number 02.207.510
Device Problem Device-Device Incompatibility (2919)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/03/2022
Event Type  malfunction  
Manufacturer Narrative
Additional device product codes: hwc, jdw, and hrs.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2022, at the time of placement of the 7.3 proximal screws, the 110 lengths of both the conical cannulated and the locked cannulated do not pass through the 2.5mm drill guide.The issue is from the rack of screws for these screws remains very narrow, and were tight inside the holes at the bottom, and this caused the thread tips to close up preventing the entry of the guides.This report involves (1) 7.3mm cannulated conical screw partially threaded/110mm.This is report 2 of 2 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
This is report 2 of 4 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h6: a manufacturing record evaluation cannot be performed due to lot number being unknown.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the cannscr ø7.3 short thread l110 sst was found to be deformed from the distal tip.Assembling issues are most likely due to this condition.A dimensional inspection for the cannscr ø7.3 short thread l110 sst and was unable to be performed due to post manufacturing damage.A functional test was unable to be performed since mating device was not returned.The complaint condition was able to be replicated.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the cannscr ø7.3 short thread l110 sst would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed -7.3mm conical screw (current) device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.B5.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
7.3MM CANNULATED CONICAL SCREW PARTIALLY THREADED/110MM
Type of Device
PIN, FIXATION, SMOOTH
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key13977010
MDR Text Key288868033
Report Number8030965-2022-02106
Device Sequence Number1
Product Code HTY
UDI-Device Identifier07611819906259
UDI-Public(01)07611819906259
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
K000066
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number02.207.510
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/03/2022
Initial Date FDA Received04/01/2022
Supplement Dates Manufacturer Received04/27/2022
06/01/2022
Supplement Dates FDA Received05/25/2022
06/01/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
CANN-LOCKSCR Ø7.3 L110 SST.; UNK - GUIDES/SLEEVES/AIMING: GUIDE.; UNK - PLATES: TRAUMA.
-
-