Model Number N/A
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/1980
Event Type  malfunction  
Manufacturer Narrative
As per the mb diagnostic catheters survey guidewire survey the respondent indicated that he had marker band movement that separated/dislodged.There was no reported patient injury.No additional information is available as this was a blinded survey.The products were not returned for analysis and a product history record (phr) review could not be performed because the lot number for this product is unknown.Without the return of the device or images for analysis, the reported customer event ¿marker band-dislodged¿ could not be confirmed.Procedural/handling factors such as entrapment of the catheter between other endovascular devices and the vessel wall or vessel characteristics may have contributed to this issue.Although not intended as a mitigation of risk, information for safety is provided in the products labeling with the intent to make the user aware of the risks.According to the instructions for use (ifu), ¿manipulation of the catheter under excessive friction due to interaction with other devices or while trapped in the vasculature, can lead to stretching or elongation of the catheter.Stretching or elongation of the catheter during endovascular procedures could result in the marker bands moving along the catheter.In extreme cases, marker bands may come off the catheter and dislodge into the vascular system.Avoid entrapment of the catheter between other endovascular devices and the vessel wall.¿ based on the information available and the phr, there is no indication that the event is related to the device design or manufacturing process.Therefore, no preventative or corrective actions will be taken at this time.
 
Event Description
As per the mb diagnostic catheters survey guidewire survey the respondent indicated that he had marker band movement that separated/dislodged.There was no reported patient injury.No additional information is available as this was a blinded survey.The device is not available for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN SUPERTORQUE CATHETER
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer (Section G)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer Contact
karla castro
miami lakes, FL 33014
7863138372
MDR Report Key13979835
MDR Text Key290462901
Report Number9616099-2022-05501
Device Sequence Number1
Product Code DQO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K922891
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberUNKSUPERTORQUEC
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/19/2022
Initial Date FDA Received04/01/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-