C.R. BARD, INC. (BASD) -3006260740 POWERPORT SLIM IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, KIT, 6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Catalog Number 1716070J
Device Problems Illegible Information (4050); Missing Information (4053)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/07/2022
Event Type  malfunction  
Manufacturer Narrative
The catalog number identified has not been cleared in the us, but is similar to the powerport slim implantable port, chronoflex single-lumen, kit, 6f products that are cleared in the us.The pro code and 510k number for the powerport slim implantable port, chronoflex single-lumen, kit, 6f products is identified.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer.However, photos were provided for review.The investigation of the reported event is currently underway.(expiry date: 04/2023).
 
Event Description
It was reported that prior to a port placement procedure, the depth markings allegedly not printed on the catheter.The procedure was completed using another device.There was no patient contact.
 
Event Description
It was reported that prior to a port placement procedure, the depth markings allegedly not printed on the catheter.The procedure was completed using another device.There was no patient contact.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us, but is similar to the powerport slim implantable port, chronoflex single-lumen, kit, 6f products that are cleared in the us.The pro code and 510k number for the powerport slim implantable port, chronoflex single-lumen, kit, 6f products is identified in d2 and g4.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one powerport slim implantable port and a catheter were returned for evaluation and two electronic photos were provided for review.Gross visual evaluation was performed on the returned device.The investigation is confirmed for the reported missing depth markings on the catheter issue as no depth marks were noted throughout the catheter was also observed on the provided photos.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 04/2023).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT SLIM IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, KIT, 6F
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key13995680
MDR Text Key288511839
Report Number3006260740-2022-01122
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741163739
UDI-Public(01)00801741163739
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K063377
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number1716070J
Device Lot NumberREFU1449
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/22/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/07/2022
Initial Date FDA Received04/04/2022
Supplement Dates Manufacturer Received04/25/2022
Supplement Dates FDA Received04/27/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/18/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-