Model Number UNK-P-AUS |
Device Problems
Fluid/Blood Leak (1250); Inflation Problem (1310); Mechanical Problem (1384)
|
Patient Problem
Urinary Incontinence (4572)
|
Event Date 03/01/2022 |
Event Type
Injury
|
Event Description
|
It was reported that, following two to three months of satisfactory performance post-implant, the patient with this artificial urinary sphincter (aus) experienced recurring incontinence due to loss of function of the device.The patient underwent surgery to remove an replace the aus.At explant, air was observed inside the aus system but no holes were noticed upon removal.There were no patient complications reported.
|
|
Manufacturer Narrative
|
Upon receipt at our quality assurance laboratory, this artificial urinary sphincter underwent a thorough analysis.The cuff was visually and microscopically examined, and leak tested.A tear consistent with a sharp instrument was identified, avulsion was present.The balloon was visually and microscopically examined, and leak tested.No anomalies were found with the balloon.The pump was visually and microscopically examined, and leak tested.No anomalies were found with the pump.Based on the information available and analysis results, the identified leak in the cuff could have caused or contributed to the reported clinical observation of mechanical issue and air in system.
|
|
Event Description
|
It was reported that, following two to three months of satisfactory performance post-implant, the patient with this artificial urinary sphincter (aus) experienced recurring incontinence due to loss of function of the device.The patient underwent surgery to remove an replace the aus.At explant, air was observed inside the aus system but no holes were noticed upon removal.There were no patient complications reported.
|
|
Search Alerts/Recalls
|
|