• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER TRAUMA KIEL LAG SCREW, STST GAMMA3® Ø10.5X90MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL LAG SCREW, STST GAMMA3® Ø10.5X90MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 40600090S
Device Problem Difficult to Remove (1528)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/25/2022
Event Type  malfunction  
Event Description
As reported: "when using the removal equipment, the operator encounters difficulties in removing the cervical screw.The cervical screw extractor handle and shaft 18066135 did not fit on the screw.The operator was finally able to remove the material, but stress was generated and the intervention time was extended.Procedure: removal of gamma nail".
 
Manufacturer Narrative
Device will not be returned.If additional information becomes available, it will be provided on a supplemental report.Clinic no longer has the device.
 
Manufacturer Narrative
Please note corrections to section d4 lot#; the lot number communicated is invalid.The reported event could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.The device inspection was not possible as the product was not returned for investigation.The device history record could not be reviewed because the affected device was not returned, and the lot number communicated is invalid.A review of the labeling did not indicate any abnormalities.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If device is returned or any further information is provided, the investigation report will be reassessed.
 
Event Description
As reported: "when using the removal equipment, the operator encounters difficulties in removing the cervical screw.The cervical screw extractor handle and shaft 18066135 did not fit on the screw.The operator was finally able to remove the material, but stress was generated and the intervention time was extended.Procedure: removal of gamma nail".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAG SCREW, STST GAMMA3® Ø10.5X90MM
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key14012738
MDR Text Key288874951
Report Number0009610622-2022-00114
Device Sequence Number1
Product Code HSB
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K043431
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 06/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number40600090S
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/11/2022
Initial Date FDA Received04/05/2022
Supplement Dates Manufacturer Received05/11/2022
Supplement Dates FDA Received06/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-