Model Number H700977 |
Device Problem
No Display/Image (1183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/23/2022 |
Event Type
malfunction
|
Event Description
|
During device preparation for the paroxysmal supraventricular tachycardia procedure, while the patient was prepped, the workmate claris was powered on and there was no display which caused a delay.The conversion cable between the display and dvi was replaced, the dvi cable was replaced, the connections were checked, the reception settings were checked, the dws was restarted and discharged, but the issue persisted.The workmate claris was replaced and the procedure was completed with no adverse patient consequences.
|
|
Manufacturer Narrative
|
One claris display workstation (dws) z440 was received for evaluation.Visual inspection revealed the front panel ports, chassis, and labels were free of physical damage.Power was applied to the dws, and the dws would not boot.The monitors would not function and no splash screen was observed.Over time the fans increased to maximum, but no other output signals happened.Based on replacement parts this symptom was isolated to the motherboard.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the investigation and information provided to abbott, the reported event was confirmed, and the root cause was attributed to the motherboard.
|
|