SMITH & NEPHEW, INC. GNS II CON INS SZ5-6 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 71420538
Device Problem Failure to Align (2522)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/11/2022
Event Type  malfunction  
Event Description
It was reported that, during total knee arthroplasty, it was found that a genesis ii insert size 5-6 11mm failed to be locked and fixed with the tibial baseplate.Surgery was resumed, after a significant delay, with a smith & nephew back-up insert.Patient was not injured as consequence of this problem.
 
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Manufacturer Narrative
The associated device was returned and evaluated.A visual inspection of the returned device does not confirm the stated failure mode.The returned device has damage among the base, more than likely from attempted insertion.A dimensional inspection was attempted, but the device was too damaged from the attempted insertion to obtain accurate measurements.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.The contribution of the device to the reported event could not be corroborated.Some potential probable causes for this event could include a fit/ sizing issue or poor insertion technique.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.Internal complaint reference number: (b)(4).
 
Manufacturer Narrative
The device was not returned for evaluation.The picture provided was reviewed and could not confirm the stated failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include a fit/ sizing issue or poor insertion technique.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GNS II CON INS SZ5-6 11MM
Type of Device
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14018423
MDR Text Key288701337
Report Number1020279-2022-01515
Device Sequence Number1
Product Code JWH
UDI-Device Identifier03596010208118
UDI-Public03596010208118
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K951987
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 05/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71420538
Device Catalogue Number71420538
Device Lot Number19DT19899
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/15/2022
Initial Date FDA Received04/05/2022
Supplement Dates Manufacturer Received05/03/2022
05/18/2022
Supplement Dates FDA Received05/06/2022
05/20/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/15/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age57 YR
Patient SexFemale
-
-