RESPIRONICS CALIFORNIA, LLC RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/14/2022 |
Event Type
malfunction
|
Event Description
|
It was reported to philips the numbers do not match the flow sheet.The device was in clinical use at the time the issue was discovered, there was no patient or user harm reported.The customer has a v60 and claims it changed to default settings.The customer was unable to find any error codes in the significate log.Philips remote service engineer spoke with the customer, and stated whenever you change a mode, the settings and alarm limits do not follow.The user would need to make the appropriate changes.Further information has been requested.If additional information becomes available at a later date, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
Many good faith efforts have been made to obtain additional information from the customer however there was no response.The resolution and disposition are unknown.Based on information provided and/or service performed, the customers alleged malfunction was not confirmed.
|
|