Model Number CLV-S190 |
Device Problem
Insufficient Cooling (1130)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/09/2022 |
Event Type
malfunction
|
Event Description
|
The customer reported to olympus during preparation for an endoscopy procedure the device gave a light source error (e101) and after troubleshooting it was discovered the light source ventilation fan was not working.Customer switched to another system to complete the procedure.There was no patient harm or user injury reported due to the event.
|
|
Manufacturer Narrative
|
The evaluation confirmed the reported issue of error message and ventilation fan not working.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.It has been over 9 years since the subject device was manufactured.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the legal manufacturer's investigation, it was determined that e101 occurred due to the increased temperature of the light source because the fan stopped working.Olympus will continue to monitor field performance for this device.
|
|