The eli 380 device is indicated for use to acquire, analyze, display, and print electrocardiograms.Device is indicated for use to provide interpretation of the data for consideration by a physician.Device is indicated for use in a clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician.It is not intended as a sole means of diagnosis.The interpretations of ecg offered by the device are only significant when used in conjunction with a physician over-read as well as consideration of all other relevant patient data.Device is indicated for use on adult and pediatric populations.The device is not intended to be used as a vital signs physiological monitor.If a monitor is not readable or difficult to read the results the user would question the integrity of the device and would likely remove it from clinical use until it was repaired or replaced.If the eli 380 device loses connection to the wireless network, the device is unable to transmit or receive ekg¿s.Based on an analysis of the risk file, a failed connection on an eli380 may result in a delay of patient treatment which may cause or contribute to a serious injury or death.Therefore, hillrom is reporting this alleged connection failure as a product malfunction.The hillrom technician found the cable between the display and motherboard to be the cause of the display issue and the cable was replaced to correct the reported issue.The connectivity issue was found to be related to the pcba mainboard being corrupted.This was also replaced to resolve the issue.Based on this information, no further action is required at this time.
|