Model Number SERV RELI380-P-D
Device Problems Display Difficult to Read (1181); Intermittent Communication Failure (4038)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/26/2021
Event Type  malfunction  
Event Description
The customer reported that their device had display issues and wi-fi connection issues.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # c-(b)(4).
 
Manufacturer Narrative
The eli 380 device is indicated for use to acquire, analyze, display, and print electrocardiograms.Device is indicated for use to provide interpretation of the data for consideration by a physician.Device is indicated for use in a clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician.It is not intended as a sole means of diagnosis.The interpretations of ecg offered by the device are only significant when used in conjunction with a physician over-read as well as consideration of all other relevant patient data.Device is indicated for use on adult and pediatric populations.The device is not intended to be used as a vital signs physiological monitor.If a monitor is not readable or difficult to read the results the user would question the integrity of the device and would likely remove it from clinical use until it was repaired or replaced.If the eli 380 device loses connection to the wireless network, the device is unable to transmit or receive ekg¿s.Based on an analysis of the risk file, a failed connection on an eli380 may result in a delay of patient treatment which may cause or contribute to a serious injury or death.Therefore, hillrom is reporting this alleged connection failure as a product malfunction.The hillrom technician found the cable between the display and motherboard to be the cause of the display issue and the cable was replaced to correct the reported issue.The connectivity issue was found to be related to the pcba mainboard being corrupted.This was also replaced to resolve the issue.Based on this information, no further action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RECON ELI 380 TOUCH ASSY W/ENGLISH KEYBD
Type of Device
ELECTROCARDIOGRAPH
Manufacturer (Section D)
MORTARA INSTRUMENT
7865 n 86th street
milwaukee WI 53224
Manufacturer Contact
antonio acquafredda
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key14039939
MDR Text Key288813270
Report Number2183461-2022-00020
Device Sequence Number1
Product Code DPS
UDI-Device Identifier00732094273885
UDI-Public732094273885
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142105
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberSERV RELI380-P-D
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/26/2021
Initial Date FDA Received04/07/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-