Model Number 10-85-00
Device Problem Failure to Auto Stop (2938)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/08/2022
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.(b)(4) manufactures the s5 double head pump.The incident occurred in (b)(6).Analysis of pump serial read-out revealed an error message stored in the micro-controller which confirmed the reported event.The reported error code can be due to: defective hall sensor in the pump head; defective motor control board.A livanova field service representative was dispatched to the facility to investigate the device and could not reproduce the reported issue.However, the motor control board was replaced as precaution.The most likely root cause is an intermittent failure of the motor control board.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova received report that during the ecc, while administering cardioplegia dose, a s5 double head pump gave motor control failure error message.Reportedly, the pump did not stop but the customer improved the issue by turning the pump off and on.No further issues occurred during the surgery.The surgery was completed without problems.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 DOUBLE HEAD PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM  80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key14043302
MDR Text Key288802844
Report Number9611109-2022-00161
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 04/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-85-00
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/09/2022
Initial Date FDA Received04/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/30/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-