LIVANOVA DEUTSCHLAND S5 DOUBLE HEAD PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number 10-85-00 |
Device Problem
Failure to Auto Stop (2938)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/08/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.(b)(4) manufactures the s5 double head pump.The incident occurred in (b)(6).Analysis of pump serial read-out revealed an error message stored in the micro-controller which confirmed the reported event.The reported error code can be due to: defective hall sensor in the pump head; defective motor control board.A livanova field service representative was dispatched to the facility to investigate the device and could not reproduce the reported issue.However, the motor control board was replaced as precaution.The most likely root cause is an intermittent failure of the motor control board.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova received report that during the ecc, while administering cardioplegia dose, a s5 double head pump gave motor control failure error message.Reportedly, the pump did not stop but the customer improved the issue by turning the pump off and on.No further issues occurred during the surgery.The surgery was completed without problems.There was no report of patient injury.
|
|