• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR 20/30 PRIORITY PACK INDEFLATOR; ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR 20/30 PRIORITY PACK INDEFLATOR; ACCESSORIES Back to Search Results
Catalog Number UNKNOWN INDEFLATOR
Device Problem Difficult to Remove (1528)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2022
Event Type  malfunction  
Manufacturer Narrative
Event date: estimated date of event.The udi number is not known as the part and lot number were not provided.The device was not returned for analysis.A review of the lot history record of the reported lot could not be conducted because the part and lot number was not provided.It is possible that during guide wire removal the seal was not fully opened resulting in the reported difficult to remove; however as the device was not returned for analysis the investigation determined a conclusive cause for the reported difficulties cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event Description
It was reported that a guide wire was stuck in the copilot so the guide wire was removed along with the copilot.There were no reported adverse patient effects and there was no reported clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
20/30 PRIORITY PACK INDEFLATOR
Type of Device
ACCESSORIES
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14043796
MDR Text Key288800456
Report Number2024168-2022-03679
Device Sequence Number1
Product Code MAV
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K961471
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN INDEFLATOR
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/14/2022
Initial Date FDA Received04/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-