Model Number CLV-S190
Device Problem Insufficient Cooling (1130)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The suspect device was returned to olympus and evaluated.The evaluation confirmed the reported problem and determined that the fan motor failed.In addition, the evaluation found the scope detection sensor was worn, the harness of the emergency lamp was noted stripped and the remaining amount of gas in the lamp was noted low.The investigation is ongoing and the definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Event Description
A user facility reported to olympus that the cooling fan malfunctioned.There was no patient injury, associated with the problem, reported to olympus.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and to correct information provided on the initial report.The following section was corrected: h8.It has been over 9 years since the subject device was manufactured.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the legal manufacturer's investigation, it was determined that the cause of the phenomenon was the failure of the cooling fan.Since no further information was obtained, they were unable to determine the cause of the cooling fan failure.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA ELITE XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14047822
MDR Text Key298466148
Report Number8010047-2022-05918
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170301995
UDI-Public04953170301995
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 06/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-S190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/15/2022
Initial Date FDA Received04/07/2022
Supplement Dates Manufacturer Received05/13/2022
Supplement Dates FDA Received06/06/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/13/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-