Model Number CLV-S190 |
Device Problem
Insufficient Cooling (1130)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The suspect device was returned to olympus and evaluated.The evaluation confirmed the reported problem and determined that the fan motor failed.In addition, the evaluation found the scope detection sensor was worn, the harness of the emergency lamp was noted stripped and the remaining amount of gas in the lamp was noted low.The investigation is ongoing and the definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
|
|
Event Description
|
A user facility reported to olympus that the cooling fan malfunctioned.There was no patient injury, associated with the problem, reported to olympus.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and to correct information provided on the initial report.The following section was corrected: h8.It has been over 9 years since the subject device was manufactured.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the legal manufacturer's investigation, it was determined that the cause of the phenomenon was the failure of the cooling fan.Since no further information was obtained, they were unable to determine the cause of the cooling fan failure.Olympus will continue to monitor field performance for this device.
|
|