The company service representative examined the system and replicated the reported event.To resolve the issue, the nonconforming laser indirect ophthalmoscope (lio) laser fiber cable was replaced.The system was then tested and met all product specifications.A system manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this lot/batch/serial number was performed.No similar complaints were reported for the product lot/batch/serial under investigation.The root cause of the reported event is attributed to nonconforming laser indirect ophthalmoscope (lio) laser fiber cable.However, the ophthalmic system itself was found to meet specifications.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|