• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 SENSICA SURESTEP TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 SENSICA SURESTEP TRAY Back to Search Results
Model Number SCCS1002
Device Problems Nonstandard Device (1420); Device Sensing Problem (2917)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/14/2022
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that during an evaluation, the sensica system with a119108s would not recognize the pre-printed tube band in order to advance to the calibration screen.Mss tried it multiple different times but could not get it to work.When the mss used the tube band that came with the ring, the system recognized it right away.The device was not used on a patient, but used only in training.
 
Manufacturer Narrative
Per investigator evaluation, bd has determined that this mdr event is not reportable.H11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the device was not returned.
 
Event Description
It was reported that during an evaluation, the sensica system with a119108s would not recognize the pre-printed tube band in order to advance to the calibration screen.Mss tried it multiple different times but could not get it to work.When the mss used the tube band that came with the ring, the system recognized it right away.The device was not used on a patient, but used only in training.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSICA SURESTEP TRAY
Type of Device
SENSICA SURESTEP TRAY
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key14054247
MDR Text Key289020673
Report Number1018233-2022-02149
Device Sequence Number1
Product Code EXS
UDI-Device Identifier00801741215773
UDI-Public00801741215773
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSCCS1002
Device Catalogue NumberA119108S
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/14/2022
Initial Date FDA Received04/08/2022
Supplement Dates Manufacturer Received09/22/2022
Supplement Dates FDA Received09/27/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-