• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION VALVE SET, EM2400; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION VALVE SET, EM2400; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938724
Device Problems Fitting Problem (2183); Free or Unrestricted Flow (2945)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that there was a connection issue with an unspecified quantity of em2400 valve sets.The events were further described as the ¿inlets are not occluding the lipid port when it is not in use and allowing some lipid to be pulled into fluids in which they don't belong.¿ it was further reported that when they tried to configure the tube set and bypass the lipid ports, it was just pulling air causing air in line issues.This issue occurred during compounding when running bags that were supposed to be clear normal saline (ns) and dextrose 5% in water (d5w) which appeared murky and cloudy from lipid contamination.As a result, cloudy solutions were discarded.Tpns (total parenteral nutrition) with lipids were in range so they were dispensed.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: the actual devices were not available; however, two (2) companion samples were received for evaluation.Unaided visual inspection did not identify any abnormalities that could have contributed to the reported condition.System level testing was performed on both samples using an in-lab compounder and no leaks were observed; accuracy testing met specifications.The reported condition was not verified on the companion samples.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALVE SET, EM2400
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
AVAILMED
c. industrial lt. 001 mz. 105
no 20905 int a, col cd ind.
tijuana, baja california 22444
MX   22444
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key14057694
MDR Text Key288909516
Report Number1416980-2022-01686
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477183
UDI-Public(01)00085412477183
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K002705
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial,Followup,Followup
Report Date 10/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberH938724
Device Lot Number60326274
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/11/2022
Initial Date FDA Received04/08/2022
Supplement Dates Manufacturer Received05/05/2022
10/21/2022
Supplement Dates FDA Received05/09/2022
10/25/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/19/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
COMPOUNDER DEVICE; D5W; NORMAL SALINE; TPN
-
-