The catalog number identified in has not been cleared in the us, but is similar to the powerport implantable port products that are cleared in the us.The pro code and 510k number for the powerport implantable port products is identified.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, a video was provided for review.The investigation of the reported event is currently underway.(expiration date: 03/2021).Device not returned.
|
H10: the catalog number identified in section d4 has not been cleared in the us, but is similar to the powerport implantable port products that are cleared in the us.The pro code and 510k number for the powerport implantable port products is identified in d2 and g4.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical sample was not returned for evaluation, one electronic video was provided for review.The video shows one mri powerport attached to a catheter.The clinician is trying to infuse the port using a infusion needle.Fluid leak was noted from the catheter near the proximal end of the catheter.Therefore, the investigation is confirmed for the reported fluid leak and fracture issues.However, the investigation is inconclusive for the reported restricted flow rate as the provided video is not sufficient enough to confirm the issue and further the exact circumstances at the time of the reported event cannot be confirmed.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiration date: 03/2021), h11: h6 (result, conclusion), h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|