• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II VIDEO SYSTEM CENTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II VIDEO SYSTEM CENTER Back to Search Results
Model Number CV-180
Device Problems No Display/Image (1183); Image Display Error/Artifact (1304)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/13/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(6).The evaluation confirmed the reported issue.Additionally, the concerned product has been repaired twice in the past year.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
 
Event Description
The customer reported to olympus device had no image, only stripes.There was no death or injury reported due to the event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.It has been over 14 years since the subject device was manufactured.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the legal manufacturer's investigation, the phenomenon of ¿no image¿ was reproduced and was determined to be due to faulty dp board.The phenomenon of ¿color bars appeared¿ was not reproduced.The cause of failure could not be identified.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II VIDEO SYSTEM CENTER
Type of Device
VIDEO SYSTEM CENTER
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14074079
MDR Text Key298596091
Report Number8010047-2022-06045
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170215513
UDI-Public04953170215513
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCV-180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/13/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/13/2022
Initial Date FDA Received04/11/2022
Supplement Dates Manufacturer Received05/06/2022
Supplement Dates FDA Received05/25/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/08/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-