Model Number CV-180 |
Device Problems
No Display/Image (1183); Image Display Error/Artifact (1304)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/13/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).The evaluation confirmed the reported issue.Additionally, the concerned product has been repaired twice in the past year.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
|
|
Event Description
|
The customer reported to olympus device had no image, only stripes.There was no death or injury reported due to the event.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.It has been over 14 years since the subject device was manufactured.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the legal manufacturer's investigation, the phenomenon of ¿no image¿ was reproduced and was determined to be due to faulty dp board.The phenomenon of ¿color bars appeared¿ was not reproduced.The cause of failure could not be identified.Olympus will continue to monitor field performance for this device.
|
|