• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NOVA BIOMEDICAL CORP. NOVA STATSTRIP GLUCOSE HOSPITAL METER SYSTEM; GLUCOSE TEST SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NOVA BIOMEDICAL CORP. NOVA STATSTRIP GLUCOSE HOSPITAL METER SYSTEM; GLUCOSE TEST SYSTEM Back to Search Results
Model Number GLU TEST STRIP
Device Problem Failure to Sense (1559)
Patient Problem Insufficient Information (4580)
Event Date 10/21/2021
Event Type  malfunction  
Manufacturer Narrative
There is currently a pending investigation.Nova is requesting additional information, and further details will be provided in a supplemental report.
 
Event Description
Nova biomedical (nova) was made aware of a potential issue regarding analysis of glucose results during the treatment of a patient while using an unidentified nova statstrip glucose meter.The user indicated the meter was asking for a sample to be applied to the disposable test strip of lot 0321069249, but it did not recognize when the sample was applied.The meter recognized the strip when inserted but did not initiate analysis when the sample was applied.Another test strip was immediately used which provided a result.No patient harm or adverse treatment due to the lack of results was reported.
 
Manufacturer Narrative
Udi:(b)(4).No product was returned to nova biomedical for investigation.The manufacturer retains of the test strip lot 0321069249 were tested using five retained meters.The test strips were used during analysis of multiple samples.During this usage, the reported error was not seen with the retained strips and the reported problem could not be reproduced.Device history record (dhr) reviews were performed for the test strips being used at the time of the event by a quality control investigator.The reviews included an assessment of the production, testing, and release of the meter and test strips.No abnormalities or concerns were noted, and the dhr indicated the released product met all specifications.The conclusion of the investigation is the reported customer complaint could not be reproduced after testing the retained test strips.A conclusive root cause was unable to be identified, and nova will continue to monitor for recurrence of similar events.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOVA STATSTRIP GLUCOSE HOSPITAL METER SYSTEM
Type of Device
GLUCOSE TEST SYSTEM
Manufacturer (Section D)
NOVA BIOMEDICAL CORP.
200 prospect st
waltham MA 02454 1941
Manufacturer (Section G)
NOVA BIOMEDICAL CORP.
200 prospect st
waltham MA 02454 1941
Manufacturer Contact
samantha cox
200 prospect st
waltham, MA 02454-1941
7816473700
MDR Report Key14077781
MDR Text Key289102253
Report Number1219029-2022-00021
Device Sequence Number1
Product Code CGA
UDI-Device Identifier10385480422144
UDI-Public10385480422144
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K063821
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/10/2023
Device Model NumberGLU TEST STRIP
Device Catalogue Number42214
Device Lot Number0321069249
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/17/2022
Initial Date FDA Received04/11/2022
Supplement Dates Manufacturer Received04/11/2022
Supplement Dates FDA Received05/05/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/10/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-