Model Number N/A
Device Problems Difficult to Insert (1316); Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/02/2022
Event Type  malfunction  
Event Description
It was reported the outer cup and liner of the bipolar cup could not be assembled because the locking ring was bent.The product could not be used.No additional information.
 
Manufacturer Narrative
(b)(4).Report source: foreign: country: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.An e1 ringloc bipolar 28x51mm was returned and evaluated against the complaint.Visual inspection found the liner and cup to be in pristine condition.No scratching or damage to either component was observed.The ring is bent.Comparing the bend in the ring to the photos provided in the subform, the ring in the photo is confirmed to be oriented incorrectly with flat part of the ring facing outward.The chamfer is facing towards the inner radius of the cup.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
E1 RINGLOC BIPOLAR 28X51MM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key14078690
MDR Text Key289108854
Report Number0001825034-2022-00869
Device Sequence Number1
Product Code JDI
UDI-Device Identifier00880304568785
UDI-Public(01)00880304568785(17)260401(10)278680
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K051569
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number110010462
Device Lot Number278680
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/21/2022
Initial Date FDA Received04/11/2022
Supplement Dates Manufacturer Received05/05/2022
Supplement Dates FDA Received05/10/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-