Model Number N/A |
Device Problems
Difficult to Insert (1316); Material Twisted/Bent (2981)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/02/2022 |
Event Type
malfunction
|
Event Description
|
It was reported the outer cup and liner of the bipolar cup could not be assembled because the locking ring was bent.The product could not be used.No additional information.
|
|
Manufacturer Narrative
|
(b)(4).Report source: foreign: country: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.An e1 ringloc bipolar 28x51mm was returned and evaluated against the complaint.Visual inspection found the liner and cup to be in pristine condition.No scratching or damage to either component was observed.The ring is bent.Comparing the bend in the ring to the photos provided in the subform, the ring in the photo is confirmed to be oriented incorrectly with flat part of the ring facing outward.The chamfer is facing towards the inner radius of the cup.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|