• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS AG SL-PLUS STANDARD STEM 5 NON-CEMENTED; PROSTHESIS, HIP, HEMI-, FEMORAL, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS AG SL-PLUS STANDARD STEM 5 NON-CEMENTED; PROSTHESIS, HIP, HEMI-, FEMORAL, METAL/POLYMER Back to Search Results
Model Number 75002703
Device Problem Loss of Osseointegration (2408)
Patient Problems Joint Dislocation (2374); Inadequate Osseointegration (2646)
Event Date 05/15/2014
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: case (b)(4).
 
Event Description
It was reported that, after a left tha had been performed on (b)(6) 2008, the patient experienced loosening of the femoral stem and a joint dislocation.A revision surgery was performed on (b)(6) 2014 to treat this adverse event.During this procedure, all the tha prosthesis construct was explanted and replaced with an unspecified system.The patient's health status is not known.This information was provided by the (b)(6), following a review of the data conducted in september 2021.
 
Manufacturer Narrative
H3, h6: a total of 84 revision surgeries involving bicon-plus devices was reported by the uk national joint registry.140 complaints were opened accordingly.A review of the device history record (dhr), of the complaint history (chr), of the instruction for use (ifu), of the risk management file (rmr) and of past escalation (per) was performed.No relevant supporting clinical information was provided.Based on insufficient information, a thorough clinical assessment cannot be performed at this time.The patient's current condition is unknown and the patient impact beyond the reported events could not be determined.Based on the chr and dhr evaluation performed there is no indication that any failure to meet specification contributed to the reported events.The reported failure modes are outlined in the device instruction for use lit no 12.23 03/21.The occurrence and severity of the reported issues are in line with the risk plans.The root cause of the reported events was attributed to known inherent risks of the devices.No corrective actions have been initiated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SL-PLUS STANDARD STEM 5 NON-CEMENTED
Type of Device
PROSTHESIS, HIP, HEMI-, FEMORAL, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ   CH-5000
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14079994
MDR Text Key289062024
Report Number9613369-2022-00083
Device Sequence Number1
Product Code KWY
UDI-Device Identifier07611996078671
UDI-Public07611996078671
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K072852
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/28/2015
Device Model Number75002703
Device Catalogue Number75002703
Device Lot NumberF0804066
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/22/2022
Initial Date FDA Received04/12/2022
Supplement Dates Manufacturer Received05/11/2022
Supplement Dates FDA Received05/30/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/29/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age56 YR
Patient SexFemale
-
-