Model Number BIPAP FOCUS SYSTEM, INT'L, RENTAL |
Device Problem
Power Problem (3010)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/22/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that the device will power on but does not indicate ac when connected.The device then goes ventilator inoperable with a 437 error code.There was no patient involvement when the issue occurred, no patient or user harm reported.Per the diagnostics performed by the engineer, code 437 involves a battery voltage failure.The customer was not getting any power out of the power supply.The remote service engineer (rse) provided the part number for the power supply.The customer reported that the power supply was replaced, and the issue was resolved.The device was returned to service.
|
|