Model Number M00510880
Device Problems Break (1069); Separation Failure (2547)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/14/2022
Event Type  malfunction  
Manufacturer Narrative
Initial reporter facility address: (b)(6).(b)(4).A supplement mdr is not warranted; the device was not returned, the investigation results do not alter the original conclusions in the initial report, and there is no new information that would prompt bsc to alter any information previously submitted.
 
Event Description
It was reported to boston scientific corporation that a trapezoid rx basket was used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2022.During the procedure, a trapezoid rx basket was used in conjunction with an alliance handle in an attempt to crush a stone about 2 cm.However, the basket failed to crush the stone and the tip failed to separate.The stone was stuck inside the basket and then the handle broke.The basket with the stone was pulled out from the patient.A different device was used to complete the procedure.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAPEZOID RX
Type of Device
LITHOTRIPTOR, BILIARY MECHANICAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key14091225
MDR Text Key289319908
Report Number3005099803-2022-01849
Device Sequence Number1
Product Code LQC
UDI-Device Identifier08714729296393
UDI-Public08714729296393
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K040447
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/02/2022
Device Model NumberM00510880
Device Catalogue Number08714729296393
Device Lot Number0028485957
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/16/2022
Initial Date FDA Received04/12/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/02/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age84 YR
Patient SexFemale
Patient Weight41 KG
-
-