The subject device was returned to olympus for further evaluation.Inspection and testing found there was abnormal insufflation due to disconnection of the solenoid valve interface connection wire.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device met all specifications at the time of shipment.Based on the results of the legal manufacturer's investigation, it is likely the air pressure was incorrect due to disconnection of the solenoid valve interface connection wire.However, the root cause of the wire disconnection could not be determined.Olympus will continue to monitor field performance for this device.This late mdr is being submitted as a result of a complaint retrospective review.
|