Catalog Number 11012-39
Device Problems Entrapment of Device (1212); Difficult to Remove (1528); Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/22/2022
Event Type  Injury  
Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that the procedure was to treat the iliac artery with moderate calcification and moderate tortuosity.The 9x39 mm omnilink elite stent delivery system (sds) was prepared per the instructions for use (ifu).During advancement the sds became stuck in the calcified lesion.The sds was not able to be removed and could not cross the lesion so bypass surgery was performed to remove the sds and complete the procedure.The patient was hospitalized extra time due to the issue.No additional information was provided.
 
Event Description
N/a.
 
Manufacturer Narrative
A visual inspection was performed on the returned device.The reported failure to advance and entrapment of the device could not be tested as they were based on operational circumstances.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents.The investigation determined the reported difficulties in addition to removal of foreign body and surgical intervention appear to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design, or labeling.H10- removed medical device problem code 1528; added 1212.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNILINK ELITE PERIPHERAL STENT SYSTEM
Type of Device
PERIPHERAL STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14106875
MDR Text Key289257472
Report Number2024168-2022-04024
Device Sequence Number1
Product Code NIO
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P110043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number11012-39
Device Lot Number1070841
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/22/2022
Initial Date FDA Received04/13/2022
Supplement Dates Manufacturer Received06/29/2022
Supplement Dates FDA Received07/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/08/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
-
-