Model Number V173
Device Problems Over-Sensing (1438); Pacing Problem (1439); Pocket Stimulation (1463); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problems Bradycardia (1751); Device Overstimulation of Tissue (1991); Dizziness (2194)
Event Date 03/23/2022
Event Type  Injury  
Event Description
It was reported that the patient with this cardiac resynchronization therapy pacemaker (crt-p) presented to the emergency room with dizziness and pocket stimulation.The crt-p had entered in safety mode, and was explanted due to oversensing, and pacing inhibition with asystole less than 2 seconds.Another device was successfully replaced.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INVIVE
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key14107839
MDR Text Key289250046
Report Number2124215-2022-10133
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526536632
UDI-Public00802526536632
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S079
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/17/2017
Device Model NumberV173
Device Catalogue NumberV173
Device Lot Number227449
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/24/2022
Initial Date FDA Received04/13/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/17/2015
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age72 YR
Patient SexMale
-
-