Model Number MESE1 |
Device Problems
Difficult or Delayed Activation (2577); Ventilation Problem in Device Environment (3027)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/01/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Date of event: unknown, assumed 1st day of month that the event took place.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Serial number was received and dhr is pending review.When the review is completed, a supplemental medwatch will be sent with a summary of the evaluation.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
It was reported during an unknown procedure that the unit reads rf activation port not recognizing.Unable to test.The unit continues to run or does not run at all.
|
|
Manufacturer Narrative
|
(b)(4).Date sent; 5/3/2022 investigation summary per service manual operational and diagnostic analysis confirmed the unit will not run.The scroll pump and pcb assembly board were replaced as identified in the investigation to address the issue.The repair and testing of the unit was completed per the service manual, bringing the unit back to full functionality.The unit passed all functional tests and is fully operational.The repair and testing of the unit was completed.The device history records were reviewed and certed by external manufacturing that the manufacturing criteria were met prior to the release of the equipment.The certificate records are accessible through external manufacturing.
|
|