It was reported to aesculap inc.That a prestige atra grasper dbl-act 5mm (part # 8360-10) was used during a procedure performed on (b)(6) 2022.According to the complainant, during the procedure, the screw holding the distal tip together detached.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of this event.Although requested, additional information has not been made available.The malfunction is filed under aag reference (b)(4).
|
Manufacturer evaluation: the device was returned to the manufacturer for physical evaluation.A visual examination of the device was performed.Which revealed, that the arms of the actuator appeared to have flared open causing disassembly at the distal tip.No evidence of any screws missing or being an issue (no screws/threads in this location of the device).Both jaws were returned.However, only one link was returned.The other link was not provided.An investigation of the device manufacturing records was not able to be conducted by the manufacturer, as the lot # was not made available.However, all device history records (dhr) are reviewed and released according to documented procedures.And a device is not released if it does not meet requirements or is nonconforming.The investigation into the cause of the reported problem was able to confirm, the failure mode of distal tip failure.In addition, to a supplier corrective action request (scar) being initiated, a corrective action/preventive action (capa) was opened by aesculap inc.For further evaluation of the design transfer of this device.
|