A discordant, falsely low free light chains, type lambda (flc lambda) result was obtained on a patient sample on a bn ii system (serial number: (b)(4) ) with n latex flc lambda reagent (lot number: 473264) using a 1:5 sample dilution.The sample was then repeated for flc lambda on an alternate bn ii system (serial number: (b)(4) ) with an alternate lot of n latex flc lambda reagent (lot number: 473263) using a 1:2000 sample dilution, recovering falsely low.The discordant results were not reported to the physician(s).The sample was then sent to an alternate laboratory and was run for flc lambda on a bn ii system with a non-siemens reagent using a 1:8000 sample dilution, recovering higher.This higher result was reported, as the correct result, to the physician(s).A discordant, falsely elevated free light chains, type kappa (flc kappa) result was obtained on the same sample on a bn ii system (serial number: 104495) with n latex flc kappa reagent (lot number: 473164) using a 1:100 sample dilution.The discordant result was not reported to the physician(s).The sample was sent to an alternate laboratory and was run for flc kappa on a bn ii system with a non-siemens reagent using a 1:20 sample dilution, recovering lower.The lower result was reported, as the correct result, to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the discordant flc lambda and flc kappa results.
|
An outside of the united states (ous) customer contacted a siemens customer care center (ccc).Quality controls (qc) recovered in range.Siemens' investigation determined that the issue was limited to one patient sample.As per the instructions for use (ifu) for n latex flc lambda and n latex flc kappa, "excessively elevated monoclonal immunoglobulin concentrations might have the potential to suppress the reaction of anti-flc antibodies with free light chain molecules.If results do not fit to previous ones, or to results of other tests (e.G.Serum immunoelectrophoresis, immunofixation, differential blood cell count) and/or to the clinical situation, it is recommended to re-analyze the sample in a higher sample dilution.)" the patient's elevated monoclonal immunoglobulin concentrations cannot be ruled out as a cause of the discordant results.The cause of the event is unknown.The reagent is performing according to specifications.No further evaluation of this device is required.Mdr 9610806-2022-00018 was filed the discordant flc lambda result obtained with n latex flc lambda reagent (lot number: 473264).Mdr 9610806-2022-00020 was filed the discordant flc kappa result obtained with n latex flc kappa reagent (lot number: 473164).
|