The spacer 101-9812 lot# 27594003 was returned and analyzed.A labeling review found that the reported event is a known risk associated with the use of lumbar spine implants and associated instruments.However, visual inspection revealed severe abrasion on the mating surface of the actuator, and the screw thread was stripped.Although, the implant functioned acceptably during the functional test.The damage is most likely to have occurred during the explant procedure.A product labeling review identified that the device was used per the instructions for use (ifu) product label.Additionally, the following are noted within the ifu as potential risks associated with the use of the device: discomfort and rehabilitation associated with recovery from surgery, pain and discomfort associated with the operative site or presence of implants, and fracture, damage or remodeling of adjacent anatomy, including bony structures or soft tissues during or after surgery.
|