Model Number V60
Device Problems Noise, Audible (3273); Excessive Heating (4030)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/11/2022
Event Type  malfunction  
Manufacturer Narrative
Reporting institution phone number: (b)(6).
 
Event Description
The customer reported overheating and abnormal noise problem.The customer reported that the unit was not in use on patient.No reports of impact to the patient.The customer contacted product support and the users reported that the device has an overheating issue and an anomalous noise.There are no error messages.There are no reports of impact on the patient.The customer requested a check by a technician.As the device was in the covid department, the fan was sanitized.The device is pending customer approval for repair.
 
Manufacturer Narrative
The field service engineer (fse) replaced the blower and ran functional tests with a positive outcome.The ventilator is ready for clinical use.
 
Manufacturer Narrative
The blower assembly was returned for failure investigation.The blower assembly failed testing.Bad temperature sensor lm20 generated the blower temperature high error code.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer Contact
melissa abbott
2271 cosmos court
carlsbad, CA 92011
7609187300
MDR Report Key14136191
MDR Text Key289448702
Report Number2031642-2022-01008
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838009851
UDI-Public00884838009851
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053614
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/11/2022
Initial Date FDA Received04/18/2022
Supplement Dates Manufacturer Received04/22/2022
07/25/2022
Supplement Dates FDA Received05/13/2022
08/12/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/29/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse