Model Number SIGTRSB60AXT |
Device Problems
Misfire (2532); Separation Failure (2547); Device Displays Incorrect Message (2591)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/29/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during a laparoscopic sleeve gastrectomy, on the second firing on the stomach, the power handle light emitting diode (led) screen read excessive force.The surgeon waited 30 seconds and was able to continue the firing.When surgeon went to retract the knife blade and open the reload, it was noted that the knife blade had not fully reached its end mark as it did not cut the distal stitch of the two reinforced reloads and had incomplete firing.The surgeon then retracted the knife blade and cut free the buttress material from the reloads.The doctor then used a manual staple handle and new reload to resolve the issue.There was no patient injury.
|
|
Event Description
|
According to the reporter, during a laparoscopic sleeve gastrectomy, on the second firing on the stomach, the power handle light emitting diode (led) screen read excessive force.The surgeon waited 30 seconds and was able to continue the firing.When surgeon went to retract the knife blade and open the reload, it was noted that the knife blade had not fully reached its end mark as it did not cut the distal stitch of the 2 reinforced reloads and had proximal incomplete firing (a third to one half the distance of the firing).The surgeon then retracted the knife blade and cut free the buttress material from the reloads.The doctor then used a manual staple handle and new reload to resolve the issue.There was no patient injury.
|
|
Manufacturer Narrative
|
Additional information: b5, g3 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|