ABBOTT MEDICAL PENTA 3MM LEAD, 60 CM; SCS PADDLE LEAD Back to Search Results
Model Number 3228
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/05/2022
Event Type  Injury  
Manufacturer Narrative
Date of event is estimated the results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that during an ipg replacement procedure (manufacturer reference number: 3006705815-2021-06578), the physician discovered the outer coating of the lead was no longer present.The physician placed tubing around the lead to secure it.Diagnostic testing confirmed normal impedances and appropriate stimulation coverage was obtained.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTA 3MM LEAD, 60 CM
Type of Device
SCS PADDLE LEAD
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key14149517
MDR Text Key289614835
Report Number1627487-2022-02242
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05414734406246
UDI-Public05414734406246
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/28/2016
Device Model Number3228
Device Catalogue Number3228
Device Lot Number4595751
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/14/2022
Initial Date FDA Received04/19/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/05/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexFemale
Patient Weight70 KG
-
-