MAUDE Adverse Event Report: INTEGRA LIFESCIENCES MANSFIELD CERELINK ICP PROBE 1L P BOLT; ICP MICRSOSENSORS - CERELINK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES MANSFIELD CERELINK ICP PROBE 1L P BOLT; ICP MICRSOSENSORS - CERELINK Back to Search Results
Model Number 826852
Device Problems No Device Output (1435); Low Readings (2460)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/18/2022
Event Type  Injury  
Event Description
A facility reported a cerelink microsensor was inserted with icp express and after a few minutes with positive icp values, the icp went down to -50 and stop monitoring.Even switching icp express, and replacing the zero, values were not displayed.The sensor was retrieved.
 
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Manufacturer Narrative
Additional information received: the probe insertion was the 17th and the issue came out the 18th.They started monitoring the patient with an icp express, and after the failure they explanted the sensor and implanted a new one and use another icp express on (b)(6).At the moment the patient is still monitored.
 
Event Description
N/a.
 
Manufacturer Narrative
Complaint sample was not returned for evaluation (discarded) and lot number information has not been provided; therefore, an evaluation of the device could not be performed, and manufacturing records could not be reviewed.The cause(s) of the difficulty reported by the customer could not be determined; however, the possible root cause for the issue reported by the customer could be due to excessive pressure applied to product.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERELINK ICP PROBE 1L P BOLT
Type of Device
ICP MICRSOSENSORS - CERELINK
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA
Manufacturer (Section G)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
mansfield MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key14167035
MDR Text Key289744262
Report Number3013886523-2022-00182
Device Sequence Number1
Product Code GWM
UDI-Device Identifier10886704075455
UDI-Public10886704075455
Combination Product (y/n)N
PMA/PMN Number
K173192
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 07/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number826852
Device Catalogue Number826852
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/19/2022
Initial Date FDA Received04/21/2022
Supplement Dates Manufacturer Received04/22/2022
06/16/2022
Supplement Dates FDA Received05/18/2022
07/05/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-